User's Manual

Table Of Contents
1
 einnx
 Adding Measurements
InnoSight Ultrasound System
P6992-4, EN, 16/12/6
Slope
Measure the length between the intersections of two cursors Slope can be positive or negative and
measures the rate of change between the two points defined by the intersections of the cursors in cm/sec
1 Touch SlopeA crosshair cursor with horizontal and vertical axes appears on the time series window
Drag the cursor to a desired start point and release it
2 A second cursor appears Drag it to the end point then release it to complete measurement
The slope measurement is not restricted to either the horizontal or the vertical axis of the start
marker
Time
Measure the length between two vertical time lines created by two cursors The position of the horizontal
distance lines does not affect time measurements
1 Touch TimeA crosshair cursor with horizontal and vertical axes appears on the time series window
Drag the cursor to a desired start point on the horizontal axis and release it
2 A second cursor appears Drag it to the end point on the horizontal axis then release it to complete
measurement
Heart Rate
Measure the length between two vertical lines created by two cursors in beat per minute (BPM) The
position of the horizontal distance lines does not affect the heart rate
1 Touch Heart RateA crosshair cursor with horizontal and vertical axes appears on the time series
window Drag the cursor to a desired start point on the horizontal axis and release it
2 A second cursor appears Drag it to the end point on the horizontal axis then release it to complete
measurement
Measuring in Spectral Doppler Mode
Auto Trace
Trace the spectrum of Doppler waveforms
Touch High QThe trace for the Doppler spectrum will be performed automatically with measurements
calculated and displayed on the screen
Semi-Trace
Trace the Doppler spectrum with better accuracy in the measurement calculation
1 Touch Semi TraceA crosshair cursor with vertical axis appears on the time series window Drag the
cursor to a desired peak of the graph and release it
2 A second cursor appears Drag it to end diastole on the graph then release it
3 A third cursor appears Drag it to the second peak of the graph and release it
4 Touch anywhere on the screen to complete the trace
To obtain an accurate auto-trace and measurement result, a quality Spectral Doppler image with
clean noise background is required