 Adding Measurements
InnoSight Ultrasound System
P6992-4, EN, 16/12/6
When you are measuring, the indicators/lines display in yellow, allowing you to adjust as many times
as needed When you are done with measuring, use any of the following methods to complete the
measurement The indicators/lines then turns green, and the measured results (values) appear on the top
left side of the imaging screen
Two-finger tap on the scan area
Proceed with the next measurement
Touch Acquire to save the ultrasound image
To re-position the measured results, touch Result Pos
+/-
To set whether to show all or the last measured results, or to hide them all, touch Show Result
+/-
To delete measurements, touch Erase Meas to erase the last measurement added Repeat this
action, if needed, to continue erasing measurements
To directly erase all measurements, touch and hold Erase Meas
You can select whether to keep or erase the measurements added after you return to the live
scan by touching Menu > Settings > Workflow
> Auto-clear Measurement after Unfreeze
Measuring in 2D/Color/CPA Modes
Distance
Measure a distance
1 Touch DistanceA crosshair cursor appears on the image Drag the target cursor to where you want to
start measuring and release it
2 Drag the target cursor to where you want to finish measuring, then release it to complete measurement
a
Ellipse
Measure a circumference
1 Touch EllipseA crosshair cursor appears on the image Drag the target cursor to one end of the area
you want to measure and release it to set the start point of measurement
2 Drag the target cursor to the other end of the desired area, and release it to set the end point of
measurement An oval area is then measured
b
Trace
Manually trace an irregular shape
1 Touch Trace A crosshair cursor appears on the imageDrag the cursor to where you want to start
measuring and release it
2 Drag the cursor along the outline of the object to trace
Figure 28Trace an outline