 DICOM Configuration
InnoSight Ultrasound System
P6992-4, EN, 16/12/6
2 Assign a name to the server and enter it in the Name field
3 Enter the IP Address, Port and AE Title specified by your network administrator in the respective fields
4 To test the server connection, touch Test in the Echo Test section A verification message confirms a
successful connection within a few seconds
 Repeat step 1-4 to add more servers based on your needs
If you add a second storage SCP, when you export an exam or an image/video to the DICOM
server, two tasks are created in the outgoing queues
To start exams by loading patient information in a worklist, you need to configure the worklist
SCP first
Consult your network administrator about the appropriate settings
Local Host
In the Local Host section, touch Edit
The system’s default Name and IP Address, if connected successfully to the network, display in the
respective fields
In the AE Title field, enter the application entity title of your system, which identifies you as the sender and
must be unique on its network
Consult your network administrator about the appropriate settings
Managing Outgoing Queue
When you export an exam or an image/video to the DICOM server, you can monitor its status in the
outgoing queue The system provides three ways of exporting the exam For more information, see
Exporting the Exam on page 
Task status icon* Description
Upload is pending
Upload is processing
Upload is complete
Upload failed
If you export an exam as a report, the task status icon includes
In the Outgoing Queue section, touch Edit
To re-send the task to the DICOM server, touch Retry
To cancel the ongoing task, touch Stop
To clear all tasks from the outgoing queue, touch Clear all