IT Network
failure
Impact on the equipment Hazard System countermeasures
IT network
becomes
unstable
Unable to transmit exam
data to a PACS
Delay of
diagnosis
Exam data is stored in the system’s internal
storage After the IT network has resumed
stability, you can re-initiate the data transfer
Delay of transmission to a
PACS
Incorrect data
transmitted to a PACS
Mis-diagnosis
The system uses the TCP/IP and DICOM
protocols to ensure the integrity of the data
Unable to retrieve order
data from an MWL server
Delay of exam
You can initiate/create a new exam from the
system
Delay of retrieving order
data from an MWL server
Incorrect data from a
MWL server
Incorrect exam
The system uses the TCP/IP and DICOM
protocols to ensure the integrity of the data
Firewall has
broken down
Attack via network
Manipulation of
the exam data
The system closes unnecessary network
ports
Infection by computer
virus
Exam data
leakage
The system forbids installation of any
software by any user
Tale 8IT network failure recovery measures
Even when the connection to an IT network is trusted, any change of the network settings requires
immediate checkup and possible measures taken Should any of the changes below occur, perform
additional evaluation to the IT network
Changes in the network configuration (IP address, router, proxy, and so on)
Connection of additional items
Disconnection of items
Equipment update
Equipment upgrade