User manual
Rolera Thunder User Manual    26
I/O Connector 
Pinout 
The I/O (Input/Output Status) connector provides information about trigger function, DAC, and 
TTL signals. Inputs must be at least 3.15 V for a high and less than 0.9 V for a low. 
The numbers on the I/O connector diagram correspond to the numbers given to the definition of 
each of the pins. The I/O connector is a female, Hirose HR10A connector. An I/O cable (Part # 37-
513-001) to access Trigger Input (Pin 1), Frame Readout (Pin 3), Camera exposing output (Pin 2), 
and Shutter Output (Pin 4) is available. 
Figure 23. Input/Output Connector 
INPUT/OUTPUT (I/O) TABLE 
# 
I/O INPUT  DESCRIPTION 
1 
Trigger Input 
This input is internally tied high through a 4.7kΩ resistor. A rising 
edge of the Trigger Input signal initiates the trigger. The trigger 
source would normally hold this input low and then drive it high to 
initiate the trigger. 
2 
Camera Exposing Output  Active high. A high level on this output indicates that the camera is 
exposing (integrating). 
3 
Frame Readout  Active high. A high level on this output indicates that data is being 
transferred. 
4 
Shutter Output  TTL output for timing of external shutter driver. Signal is high during 
Shutter Open Delay and exposure time. The pin does not provide 
power to drive the shutter directly, so an external shutter drive 
controller is required. 
5 
Spare Input  Reserved 
6 
Trigger Ready  This output notifies accessory devices that the camera is ready to 
receive a trigger. 
7 
TX Out  Accessory Serial Transmit 
8 
RX IN  Accessory Serial Receive 
9 
Power Status  A high level on this output indicates that the camera power is 
switched on (+5 V = on, 0 V = off). 
10 
GND  System digital ground. Any external circuitry intended to interface 
with the trigger control signals must reference this ground 
connection. 
Table 4. Input/Output Definitions 