User manual
Rolera Thunder User Manual 8
2. Connect the other end of the DATA cable to the FireWire port (indicated with blue
arrow in next graphic) located on the back of the camera.
Figure 5. Rolera Thunder Rear Panel
3. Verify that the power switch on the back of the camera is in the position (|=on, O=Off).
4. Connect the power supply to the Power connector on the back of the camera.
5. Plug the power cord into the power supply and then into a suitable wall outlet.
6. Switch on the camera (power switch on the back of camera).
Adjusting the C
Mount Adapter
The Rolera Thunder features an adjustable C-mount adapter that allows you to change the focal
depth. The C-mount adapter is adjusted in the factory to nominal C-mount back focal distance.
Adjust the C-mount adapter only if you cannot achieve focus on your instrument.
If required, use a .050" hex key to adjust the adapter as follows:
1. Loosen the two setscrews and rotate the ring to the desired height (see next figure).
The amount of travel possible depends on the camera model. Tighten the screws to
lock the adapter in place, after confirming desired height.
Figure 6. Adjustable C-Mount Adapter
Travel Distance
Rolera Thunder = .215”/ .546cm
Note: A spanner wrench can be used to rotate the adapter after the setscrews have been loosened. The
spanner holes are .15" (.38 cm) dia. x .15" (.38 cm) deep and are spaced 1.25" (3.18 cm) apart. To lock the
setscrews, the travel distance must be less than .215" (.54 cm).
(2) Spanner Wrench Holes
.15" (.38 cm) dia x.15" (.38 cm) deep
1.25" (3.18 cm) apart
(2) Setscrews
(.050" hex key)