2.1 Explanation of the Sensor Code
The pH sensors are delivered with an attached sensor code which must be entered in the
Settings window. The following gives a short explanation about the information given in
the sensor code.
Example Code: SAE7 -532-205
SA: Sensor Type. Factory calibration values of the sensor types (PK7, PK8….)
E: LED Intensity of the device
7: Amplification of the measurement signal
532 205: Pre-calibration values
2.2 Conditions in the Sample
When entering the sensor settings, the Conditions in the Sample during the
measurements have to be determined. There are two important parameters to be taken
into account and which can be automatically compensated:
Temperature
Salinity
2.2.1 Temperature
Several options for Automatic Temperature Compensation of optical pH sensors are
available:
External Temperature Sensor (Pt100, temperature port)
Fixed Temperature (must be entered, kept constant and controlled)
Optical Temperature sensor connected to a channel connector (its channel number
must be selected) of a multi-channel
FireSting pro
If External Temperature Sensor or Optical Temperature Channel is selected, automatic
compensation of temperature changes on the respective pH sensor readings is activated.
The compensation temperature will be displayed in the corresponding channel row of
the main window.
FireSting pro: If External Temperature Sensor or Optical Temperature Channel is
selected, automatic compensation of temperature changes on the respective pH sensor