3.2 Preparation of Calibration Standards
For the calibration it is strongly recommended to use PyroScience pH buffer capsules
with pH 2 (item no.
PHCAL2) and pH 11 (item no. PHCAL11)
• pH 2 buffer are used for the first calibration point
• pH 11 is used for the second calibration point
For this, open the capsule by holding both ends and pulling them apart. Dissolve the
powder of the capsule in 100 mL demineralized water. Mix well and insure, that the
contents are completely dissolved.
Note: Always prepare a fresh buffer for calibration. The buffers are not
stabilized against microbial growth.
Alternatively, it is possible to perform a Custom Calibration. Prepare a buffer with
known pH value and known ionic strength. For the first calibration point the pH must be
<3 and for the second calibration point the pH must be >10. The respective pH value
and salinity must be entered in the custom calibration menu in the software.
3.3 First Calibration Point (acidic pH)
Ensure that the correct Sensor Code has been entered into the settings and a calibration
standard has been prepared correctly.