Please do NOT use commercially available buffer solutions and certified
reference materials (CRMs) used for pH electrodes. These buffers (colored and
uncolored) contain anti-microbial agents that will affect the sensor performance
irreversibly. It is important to only use PyroScience buffer capsules or self-
made buffers with known pH and ionic strength for calibration (with advice
from PyroScience).
3.1.1 Temperature
It is crucial to determine exactly the temperature of the pH calibration buffer during the
calibration process via one of the following possibilities:
• Manual adjustment of a Fixed Temperature (needs to be determined and kept
constant)
• Temperature Compensation with an External (Pt100) Temperature Sensor
connected to the temperature port of a
FireSting pro or Pico-pH, or
• Optical Temperature Sensor (calibrated) connected to a channel at a multi-channel
FireSting pro (its respective channel number needs to be entered at Optical Temp.
Channel).
For
AquapHOx, the integrated temperature sensor can be used.
3.1.2 Background Compensation
For pH Robust Screw Cap Probes, Sensor Vials, Flow-Through Cells and Sensor Spots the
fiber length needs to be entered into the software for an automatic background
compensation. This is recommended and sufficient for most applications.
Sensor Spots and Sensor Vials used in applications with low signal intensities (e.g. thick
glass) and for measurements in colored media, the option MANUAL background
compensation should be used to determine the individual luminescence background of
the sample.
For more information, please refer to the
Workbench Software Manual Chapter 6.