User Manual
Optical pH Sensors | User Manual
© PyroScience GmbH 10
3 SENSOR CALIBRATION
3.1 General Informations
The pH calibration has to be repeated periodically, depending on the individual
application (e.g. required accuracy, temperature of the sample, life-time of the sensor,
substances in sample). It is advisable to perform a 2-point calibration before every
measurement using freshly prepared buffer solutions.
The calibration of optical pH sensor differs from other measurements methods, as the
calibration points are NOT in the dynamic range of the sensor. This is due to the function
principle.
Optical pH sensors can be 2-point calibrated and pH offset adjusted.
• 1-point calibration (required): Calibration of the completely protonated sensor at
highly acidic conditions. Recommended before every measurement. pH has to be
lower than pH 3.
• 2-point calibration (highly recommended): Additional calibration of the completely
deprotonated sensor at highly basic conditions. Recommended before every
measurement. pH has to be higher than pH 10. The 2-point calibration is required
for the pH Minisensor.
• pH-offset adjustment (for complex medium): Adjustment of the turning point of the
sensor in highly complex medium (individual sample, see image below). This offset
pH has to be in the dynamic range (e.g. between pH 6.5 and 7.5 for a pH 7 sensor).