User guide

Operating menu and parameter description T82
68
Description
Use this function to display the maximum electronics temperature measured in the past
(maximum indicator).
Device temperature min.
Navigation
Diagnostics ® Measured values ® Min/max values ® Device temperature min.
Expert ® Diagnostics ® Measured values ® Min/max values ® Device temperature
min.
Description
Use this function to display the minimum electronics temperature measured in the past
(maximum indicator).
Reset device temp. min/max values
Navigation
Diagnostics ® Measured values ® Min/max values ® Reset device temp. min/max
values
Expert ® Diagnostics ® Measured values ® Min/max values ® Reset device temp.
min/max values
Description
Resets the maximum indicators for the minimum and maximum electronic temperatures
measured.
Options:
• No
• Yes
Factory settings
No
13.3.5 "Simulation" submenu
Simulation current output
Navigation
Diagnostics ® Simulation ® Simulation current output
Expert ® Diagnostics ® Simulation ® Simulation current output
Description
Use this function to switch simulation of the current output on and off. The display alternates
between the measured value and a diagnostics message of the "function check" category (C) while
simulation is in progress.
Display
Measured value display « C491 (simulation current output)
Options:
• Off
• On
Factory settings
Off