User guide

T82 Diagnostics and troubleshooting
31
Diagno
stics
number
Event text Remedial measures
Status
signal
from the
factory
Event
level from
the
factory
Changeab
le in
102 Sensor value too
high
1. Check process temperature.
2. Check sensor.
3. Check sensor type.
F Alarm
104 Backup active 1. Check electrical connection sensor 1.
2. Replace sensor 1.
3. Check connection type.
M Warning
105 Calibration interval 1. Execute calibration and reset calibration interval.
2. Switch off calibration counter.
M Warning
1)
F
106 Backup not
available
1. Check electrical connection sensor 2.
2. Replace sensor 2.
3. Check connection type.
M Warning
Diagnostics for the electronics
201 Electronic error Replace electronics. F Alarm
221 Electronic
reference
Replace electronics. F Alarm
241 Electronic software 1. Device restart.
2. Device reset.
3. Replace electronics.
F Alarm
261 Electronic modules Replace electronics. F Alarm
262 Module connection 1. Check whether the retainers and the connection of the
display module are correctly seated on the head transmitter.
2. Test the display module with other suitable head
transmitters.
3. Display module defective? Replace module.
M Warning
283 Memory content Replace electronics. F Alarm
301 Supply voltage 1. Increase supply voltage.
2. Check electrical connection for corrosion.
F Alarm
Diagnostics for the configuration
401 Factory reset Please wait during the reset process. C Warning
402 Configuration
initialization
Please wait during the initialization process C Warning
411 Up-/Download Please wait during the Up-/Download process. C Warning
431 Factory calibration Replace electronics. F Alarm
435 Linearization 1. Check configuration of sensor parameters.
2. Check configuration of special sensor linearization.
3. Contact service organisation.
4. Replace electronics.
F Alarm
437 Configuration 1. Check configuration of sensor parameters.
2. Check configuration of special sensor linearization.
3. Check transmitter settings.
4. Contact service organisation.
F Alarm
451 Data handling Please wait during the data handling process. C Warning
483 Simulation input
Deactivate simulation. C Warning
485 Simulation
measured value
491 Simulation current
output