Operator`s manual

Operator’s Manual
922259-00 Rev A 7
Full Calibration Procedure
If the probe pair has been repaired or is to be used on a DA1855A other than the
one it was originally calibrated with, the probe pair should be calibrated using
the following Full Calibration Procedure. The last step is to perform the Short
Calibration Procedure as needed for high slew rate signals.
Required Equipment
Equipment
Differential Amplifier
Oscilloscopes
Calibration Generator
DVM
DC Voltage Generator
Probe Tip to BNC Adaptor (2)
BNC T Adaptor
COAX Cables (2)
Termination, 50 Ohm
Termination, 75 Ohm
Trimmer Tool
Refer to Figure 1, The DXC100A Board Layout for the location of adjustments,
and to the schematic diagram in Figure 2, The DXC100A Schematic for guidance.
Both figures can be found at the end of this manual.