Technical information
Deltabar M PMD55
4 Endress+Hauser
Flow measurement Design and operation mode
P01-PMD55xxx-15-xx-xx-xx-001
Flow measurement with Deltabar M PMD55 and primary element, left: Orifice plate and right: Pitot tube
1Deltabar M PMD55
2 3-valve manifold
QFlow
p Differential pressure, p = p
1
– p
2
Your benefits
• Choice between five flow modes of operation:
– Volume flow
– Norm volume flow (European norm conditions)
– Standard volume flow (American standard conditions)
–Mass flow
–%
• Choice of diverse flow units with automatic unit conversion.
• Low flow cut off: when activated, this function suppresses small flows which can lead to large fluctuations
in the measured value.
• Contains two totalizers as standard. One totalizer can be reset to zero.
• The totalizing unit can be individually set for each totalizer. This allows independent daily and annual
quantity totalizing.
Note!
For more information about the Deltatop flow measurement system, see
• TI00422P: Deltatop Differential Pressure Flow Measurement with Orifices
• TI00425P: Deltatop Differential Pressure Flow Measurement with Pitot Tubes
QQ
p
1
p
1
p
2
p
2
Q~
Dp
Q~
Dp
+
+
––
1
2
1
2