Specifications
Remote Operation
6005 Interface Panel Emulation 5
5-9
6005 Interface Panel Emulation
The RUSKA 7250 may be configured to emulate the IEEE-488 command set of the
RUSKA Single Channel Interface Panel (Models 6005-701 and 6005-761). See the
Interface Panel User’s Manual for a description of the protocol. The RUSKA 7250
emulation has the following differences:
1. The Calibrator is always in Remote mode (Byte 1, Bit 1).
2. All TI strip outputs must be written as OFF.
3. No special functions are implemented.
4. Any message written to the Calibrator that starts with a colon as the first character is
interpreted as a SCPI command.
To change from SCPI to Interface Panel Emulation via the remote interface, send the
following message:
:SYSTem:LANGuage “6000”.
To change from Interface Panel Emulation to SCPI via the remote interface send the
following message:
:SYSTem:LANGuage “SCPI”.
Serial Operation
The RS-232 port accepts the same SCPI commands as the IEEE-488 port. The commands
can be terminated by a carriage return (hexadecimal 0D) or a line feed (hexadecimal 0A).
The responses are always terminated by a carriage return followed by a line feed.
The serial port also supports XON/XOFF. When the XOFF (hexadecimal 13) command
is received, the RUSKA 7250 will stop transmitting. Transmission is restarted when the
XON (hexadecimal 11) command is received.
When only one unit is attached, the Control-C (hexadecimal 03) command will clear the
transmit and receive buffers and disable addressing. When addressing is disabled, the unit
will respond to commands without being addressed.
Druck DPI-510 Emulation Command Summary
Supported Commands
B <value> Set Tare amount
C0 Go to MEASURE mode.
C1 Go to CONTROL mode.
D0, D2 Transmit Pressure Reading
D1 Transmit Set-point
Fxx (Ignored)
I0 Disable SRQ
I1 SRQ on In Limit
I2 SRQ on Error
I3 SRQ on In Limit and Error
I4 SRQ on End of Conversion
I5 SRQ on Error and End of Conversion
I6 SRQ on In Limit and End of Conversion
I7 SRQ on In Limit, Error and End of Conversion
J0,J1,J2 Set Pre-programmed rate.
M Unlock keyboard
N0,N1,N2,N3,N4 Set talk mode notation
O1 Zero
P<value> Set pressure Set-point