Specifications

RUSKA 7250, 7250i, 7250xi & 7250LP
Users Manual
4-24
gkn033.bmp
Figure 4-23. Menu l Program l Run - Menu
4. Press Run [F2]. The configuration of the RUSKA 7250 stored with the program is
restored, the pressure set-point is set to the pressure value in the first step, and the
Calibrator is placed in Control mode. “Run” will now be highlighted and the program
will proceed through its steps.
5. To pause the program, press Pause [F3]. Pause will now be highlighted and the
Calibrator will continue controlling to the current set-point. The Calibrator will
maintain control of the current set-point until further instructions from the operator.
Press Continue [F4] to resume the program.
6. To stop the program, press Stop [F5]. The program will stop running but the
Calibrator will continue controlling to the current set-point.
Menu | Test
The Menu | Test section of the system is used to perform a number of the diagnostic and
tuning functions that are available with the RUSKA 7250.
gkn034.bmp
Figure 4-24. Menu | Test - Menu