Specifications
RUSKA 7250, 7250i, 7250xi & 7250LP
Users Manual
4-8
Vent
The Vent [F3] function is used to rapidly vent the pressure in the system to atmosphere.
This system does not have a designated internal vent valve. Instead, the vent mode is
implemented with a fast multi-step procedure. For gauge instruments, the Calibrator
controls pressure at the maximum rate towards zero psig. When the primary sensor reads
that the pressure is within 1% FS pressure of zero psig, the controller is turned off and the
zero solenoid valve that ties the reference port to the test port is opened. This physically
vents the test port to atmosphere. If the Calibrator is at a sub-atmospheric pressure, then
the system will control to zero psig, at maximum rate, turn off the controller, and open
the zeroing solenoid valve, again venting the test port to atmosphere. This function is not
available on permanent absolute versions of the RUSKA 7250.
Step/Jog
In addition to entering the pressure set-point through the key-pad, the pressure set-point
can be changed using either the step or the jog functions. The Step function is used
primarily when the operator is taking pressure steps in equal pressure increments. The
size of the pressure step is user defined. The Jog function is most often used when
calibrating mechanical gauges such as a dial gauge and the operator wishes to change the
pressure until the mechanical gauge indicates a cardinal point. The operator can then read
the higher resolution Calibrator to determine the actual pressure value when the
mechanical gauge is indicating a cardinal point. The Step/Jog [F5] toggles between Step
and Jog, to switch between these modes, press the function key 5 [F5].
Stepping
Assure that the word Step is highlighted in the Step/Jog [F5]. Using the rotary knob,
rotate the knob clock-wise to increase the set-point, or counter clock-wise to decrease the
controller set-point. The scratchpad will be updated to a new set-point. The step size
defaults to 10% of full scale. Press ENTER to accept the new set-point.
Setting Step Size
From the Main Menu, press Menu [F6] | Setup [F2 | User [F2]. Using the rotary knob,
move the cursor to highlight Step Size. Enter the desired step size in the current pressure
units and press ENTER.
Jogging
Assure that the word Jog is highlighted in the Step/Jog [F5]. From the Main Menu, the
pressure set-point can be modified by rotating the rotary knob clockwise to increase
pressure and counter-clockwise to decrease pressure. Each increment or decrement will
change the pressure in the least significant digit. Continually rotating the rotary knob, the
pressure will continue to change until the rotation is stopped.
Menu
The Menu key is used to access the less commonly accessed configuration type functions.
This accesses the Setup, Calibration, Program, Test and Display functions.