Specifications

RUSKA 7250, 7250i, 7250xi & 7250LP
Users Manual
B-2
Summary of Error Messages, cont.
-315 Configuration Data Lost. The configuration data has been lost. Check all
parameters to be sure they are correct.
-330 Self-Test Failed. Check the display for the test that failed.
-350 Queue Overflow. The error queue was full and messages were lost.
-400 Query Error. A read request was received when there was nothing to read.
500 Controller Malfunction. Internal control failure.
501 High Limit Exceeded. The pressure was greater than the high limit.
502 Low Limit Exceeded. The pressure was less that the low limit.
503 Slew limit Exceeded. The pressure changed faster than the slew limit
allowed.
521 Pressure Overrange. The pressure reading is outside the range of the
Calibrator.
531 Oven Temp Overrange. Either the transistor that drives the heater for the
quartz Bourdon tube sensor (Chapter 2) or the oven temperature sensor
itself is malfunctioning. To observe the oven temperature, select OK then
select MENU/DISP. Check the transistor and sensor for malfunction,
requesting service (Chapter 7) if necessary.
533 Case Pressure Overrange. Select OK, then reduce the pressure at the case
reference port to 30 psia or lower.
540 Mechanical Zeroing Needed. The zero point of the quartz Bourdon sensor is
beyond the range of the compensation circuit. The zero is adjusted by the
software but should be manually adjusted for complete accuracy.
542 Oven Control Failure. The temperature controller is unable to keep the
sensor at the proper temperature.
545 Sensor Communication Error. Unable to Communicate with RUSKA 7215xi
Sensor
546 Sensor Calibration Lost. The RUSKA 7215xi sensor has lost its calibration
and must be recalibrated.
600 Factory Data Lost. Internal factory constants have been lost. Contact Fluke
for more information.
601 Calibration Mode. The Calibrate button must be pressed before SCPI
calibration commands can be executed.
800 Solenoid Over-Temperature. The control solenoids have over-heated. Wait
until they have cooled before entering Control mode.
801 Error reading Barometric RPT Pressure Sensor
802 Control Sensor out of range
803 Controller Communication Error
804 Autotune Failed