Specifications

Troubldeshooting
Version 2011v1 49
4 Troubleshooting
See fig. 010800000.
The wheelchair has a troubleshooting diagnosis system that
can be read by the dealer with special hard and software. An
indication is also given on the controller near the ON/OFF
light (A). This can provide important information about a
malfunction.
EBAPUM-020800000-nl.doc
4.1 Fault table 'DX'
Check the malfunction/signal with the aid of the following list.
Signal Possible cause Remedy
ON/OFF light does
not go on: the
wheelchair does
not move.
The plug of the controller is not properly
placed in the 'power module'.
Place the plug properly in the 'power module'
(DX BUS).
The DX power module is behind the cover
that is on the outside under the chair against
the battery tray.
The automatic safety fuse is in function. Press the fuse. See ‘Automatic fuse’.
The batteries are not properly connected. Check the connections.
A connection of the batteries is loose. Secure the connection.
ON/OFF light
flashes 1 times.
DX module wrong.
Replace one or both modules. The "on/off"
LED indicates the condition of the entire
system. The modules all have a separate
status LED: this way, you can see which one
is faulty.
ON/OFF light
flashes 2 times.
DX accessory wrong. The CLAM or LM module displays a fault as
a result of a programming fault or short circuit
and/or overload of the accessories.
The (optional) electrical high/low adjustment
is not in the lowest position. The LED flashes
two times to indicate that the speed limitation
has been switched on: the wheelchair drives
more slowly.
A
010800000