Troubleshooting guide
21
11. Press the ENTER key
MAX SETPT X.XX MA
SPEED 100.0%
(The mA value will be equal to your input signal level)
(The speed value is the setting the MPC will use for this signal value)
12. Press the ENTER key three times to step through the speed setting value, most users will leave
this at 100.0 %, however you an change it if you wish using the UP and DOWN arrows.
13. Press the ENTER key once more to accept the calibration value
ANALOG INPUT
CALIBRATED
14. Analog input signal calibration is complete
15. Press the MENU key three times to exit back to the main operating screen.
If you receive the following message:
CURRENT DELTA
OUT OF RANGE
This indicates that there is not a wide enough range between your maximum and minimum analog
signals. The minimum signal range is 3 mA. The most likely source of this error is the user did not
change the value of the incoming analog signal when moving from the MIN calibration to the MAX
calibration.
The analog input signal calibration procedure should be performed again, ensuring that a minimum
of 3 mA difference exsists between the MIN and MAX signal levels.