Product Manual

28 29
How to Use the Analyzer Check Strip
1. Turn on the analyzer by pressing either button.
2. When INSTALL MEMO CHIP or RUN TEST is displayed, press Next until UTILITY is
displayed. Press Enter.
3. Press Enter when CHECK STRIP is displayed.
4. Hold the check strip at the base and insert the check strip, ribbed side up, into the
reflectance test strip slot when INSERT STRIP is displayed.
5. The analyzer should display PASSED, along with a checkmark icon. (If the display
reads FAILED, see the note at the end of this section.) Remove the check strip and
store it in the analyzer carrying case.
6. Press Next until EXIT is displayed. Press Enter.
7. Press Next until RUN TEST is displayed.
8. Press Enter. The analyzer is ready to run tests.
Note: If the analyzer displays FAILED:
1. Clean the CardioChek Plus analyzer test strip slot (where the check strip is
inserted into the analyzer). See Section 9, Care and Cleaning.
2. Inspect the check strip to make sure it is not dirty or damaged. Use the spare
check strip and repeat.
3. See Section 10, Troubleshooting in this user guide.
Quality Control
Controls (also known as quality control materials”) are solutions for which an expected
analyte concentration range has been established. Controls are tested to check the
performance of your test system: CardioChek Plus analyzer, MEMo Chip, and PTS Panels
test strips. Use quality control materials provided by PTS Diagnostics.
Refer to the range card provided with the controls or visit
http://www.ptsdiagnostics.com for control specications.
Healthcare professionals should follow their facility’s guidelines and policies regarding
quality assurance and the use of quality control materials.
Quality control materials should be run:
With each new shipment
With each new lot number
According to state, local, and federal regulations
Important: Check the expiration date printed on the control bottles. Do not use control
solutions that have expired.
For performing a quality control test, see the instructions below.
To perform a control test you need:
CardioChek Plus analyzer
PTS Panels test strips
Quality control materials
Quality control instructions
Quality control range card
Quality Control Testing