How to Get Help and Information Regarding Your
CardioChek Plus Analyzer
1. If the analyzer is off, press either button to turn on the analyzer. Wait for the
display to read INSTALL MEMO CHIP or INSERT STRIP.
Note: If RUN TEST is displayed, go to Step 3.
2. Press Next. The display will read RUN TEST.
3. Press Next until INFORMATION is displayed.
4. Press Enter, then TECHNICAL SUPPORT is displayed.
5. Press Enter to display contact information.
6. Press Enter, then Next for ABOUT CardioChek Plus display.
7. Press Enter to display serial number (SN) and software version information.
8. Press Next to display wireless IP address and SSID (if applicable).
9. Press Enter to return to ABOUT CardioChek Plus display.
10. Press Next until EXIT, then press Enter.
11. Press Next to return to RUN TEST.
Analyzer Check Strips
A check of the analyzer operation and optics can be performed using one of the two gray
check strips. Two are included in the analyzer carrying case. The check strip veries that
the CardioChek Plus analyzer’s electronic and optical systems are functioning properly.
To perform this verication, insert the check strip into the analyzer. The analyzer will
read the reectance of the gray check strip and indicate if the reading is within the
specied acceptable range by displaying PASSED. When the check strip is not in use,
store it in the analyzer carrying case. It is recommended that the check strip verication
be performed:
Daily
If the analyzer has been dropped
When a result is not consistent with expected results