Product Manual
24 25
How to Set the Wireless Connectivity On or O
The CardioChek Plus analyzer wireless communication has been preset to o. To turn the
wireless connectivity to on or o, follow the steps listed below:
1. If the analyzer is off, press either button to turn on the analyzer. Wait for the
display to read INSTALL MEMO CHIP or INSERT STRIP.
Note: If RUN TEST is displayed, go to Step 3.
2. Press Next. The display will read RUN TEST.
3. Press Next until SETTINGS is displayed.
4. Press Enter. WI-FI is displayed.
5. Press Enter. The display will highlight
(wireless connectivity on).
6. Press Enter to select wireless connectivity on or
press Next to highlight
(wireless connectivity off).
7. Press Enter to accept the wireless connectivity choice highlighted.
8. Press Next until EXIT is displayed.
9. Press Enter to return to SETTINGS.
10. Press Next to return to RUN TEST.
Note: If unable to run a test, wireless communication and patient ID lock may be on.
You will see a
symbol ashing. To run a test, send a new patient ID to the analyzer,
using the appropriate connectivity product. Turn o wireless communication or press
and hold both buttons on your analyzer to override the patient ID lock. See the Wireless
Connectivity Statement section for additional information regarding wireless
compatibility.
How to Set Up the CardioChek Plus Analyzer for Printing
For complete details, refer to each printer’s user guide.
Testing the Printer
1. If the analyzer is off, press either button to turn on the analyzer.
Wait for the display to read INSTALL MEMO CHIP or INSERT STRIP.
Note: If RUN TEST is displayed, go to Step 3.
2. Press Next. The display will read RUN TEST.
3. Press Next until SETTINGS is displayed.
4. Press Enter, then Next until PRINT is displayed.
5. Press Enter, then Next until TEST PRINTER is displayed.
6. Press Enter and a sample printout will be generated.
How to Print Results from Memory
Note: Up to 50 test results per chemistry and 10 control test results can be stored.
1. If the analyzer is off, press either button to turn on the analyzer. Wait for the
display to read INSTALL MEMO CHIP or INSERT STRIP.
Note: If RUN TEST is displayed, go to Step 3.
2. Press Next. The display will read RUN TEST.
3. Press Next until MEMORY is displayed, then press Enter.
4. Press Enter to select CHEM or press Next, then Enter to select CONTROL.
5. Press Next to highlight the test name, then press Enter to select (LIPIDS
for example).
6. Press Next to highlight the date/time of the test result you want to print.
7. Press Enter to print the selected results.
8. Control results can be printed by selecting CONTROL instead of CHEM and then
selecting the type of result as stated above.