Specifications

Product Reference Guide 6-9
between 5.995 and 6.005. Remove the weights and verify that the
display reads 0.00.
5. You have completed the Increasing-Load Test (Phase 1).
Shift Test (6 Pounds)
The Shift Test checks to ensure that items placed anywhere on the weigh-
ing surface of the scanner/scale ar
e weighed properly. Refer to Figure 6-2
when performing this test.
1. Place and remove, in succession, a six-pound load on the center of
eac
h of the four quadrants: 1, 2, 3, and 4, (as shown in
Figure 6-2)
of the scanner/scales weigh platter
. Verify that a reading of
between 5.995 and 6.005 lb. registers, under load, in each quad-
rant tested. Check that the display
returns to 0.00 between each
load.
2. After verifying the accuracy of each quadrant of the weighing sur-
face, remove all weight from the scale. This completes the Shift
Te
s t .
Figure 6-2. Shift Test
12
34
1
2
4
3
(Models vary)