Specifications
6-6 Magellan
TM
8500Xt Scanner
11. If the calibration was successful, the speaker sounds a single tone,
the Scale Status LED begins blinking again, and “End-” appears
in the Remote Display.
12. If the calibration was not successful, the speaker sounds five tones
indicating a scale failure, and the Scale Status LED will blink
twice, strobe (fast blinks), and then continually repeat this
sequence until reset. Remove all weight from the Weigh Platter,
and repeat the procedure starting with step
8.
13. Press the Scale Zero Button again to permanently store the cali-
bration data and exit Calibration Mode. You have completed the
calibration of the scale. The unit’s calibration must now be veri
-
fied as required by state and/or local weights and measures regula-
tions. The verification procedure must be performed to assure
that a scale will pass Weights and Measures requirements before it
is placed into commercial/retail service.
You have completed the scale calibration procedure. Before replacing the
Calibration Switch Cover and sealing the scale, you must now continue
with the calibration verification tests to complete the scale’s calibration. If
you are weighing in pounds, continue with the following procedures to
verify that the scale was successfully calibrated and that it weighs properly.
If you are weighing in kilograms, turn to the instructions titled,
.