Service manual

METTLER TOLEDO
DIVA Service Manual 35
Name: Zero the scale
Command: Z<CR>
Response: Scale is zeroed, returns scale status. (see note 4)
<LF>Shhhh<CR><ETX>
Optional Commands
Name: Request high-resolution weight
Command: H<CR>
Response: Returns decimal wt in 10x with units plus scale status. (see notes 2
and 3)
<LF>xxx.xxxUU<CR><LF>Shhhh<CR><ETX>
Returns lb-oz wt in10x with units plus scale status.
<LF>xLB<SP>xx.xxOZ<CR><LF>Shhhh<CR><ETX>
Scale status only if wt < 0, initial zero error, in motion or out of capacity.
<LF>Shhhh<CR><ETX>
Name: Change units of measure
Command: U<CR>
Response: Changes units of measure, returns new units and scale status.
<LF>uu<CR><LF>Shhhh<CR><ETX>
Name: Request metrology raw counts
Command: M<CR>
Response: Returns normalized raw counts and scale status.
<LF>xxxxxxMM<CR><LF>Shhhh<CR><ETX>
Name: unrecognized command
Command: all others
Response: Unrecognized command
<LF>?<CR><ETX>
NOTES:
1) Weight field is always six characters (5 for weight, 1 for decimal point), regardless
of display size.
2) Leading zeroes are not suppressed.
3) High resolution weight field is seven characters (6 for weight, 1 for decimal point),
regardless of display size.
4) If scale is outside zero range or in motion the ā€˜Z’ command is ignored and scale
status is returned without zero status flag set, i.e., scale not at zero.