Service manual

METTLER TOLEDO
DIVA Service Manual 20
Note: If the scale is unable to complete calibration it will display “E 20”. Cycle
power to the scale and restart at Step 1 above.
12. The scale will now save the calibration and cycle out of the Service mode and return to normal
weighing mode.
13. Verify the calibration was successful with the calibrated weight. If for some reason the
calibration was not successful, restart at Step 1.
14. If applicable, you should remove power from the scale and apply the appropriate calibration
seal.
15. Place the scale into service!
Calibration Quick Reference Chart
1. Break the calibration seal and press the calibration button.
2. Select “Yes” in Group 2, Step 6 of the Service Mode.
3. Empty the scale and take the no load reading.
4. Place full load on the scale and take the reading.
5. Save the settings and verify the calibration in normal weighing mode.