Service manual

METTLER TOLEDO
DIVA Service Manual 10
3 Set-up and Calibration
There are two types of setup modes. The first is called “Service Mode” and it allows full access to all
soft switches and to calibrate the scale. The second is called “Master Mode” and it only allows
access to soft switches that do not affect the metrology of the scale.
3.1 Service setup mode:
In order to access the Service Mode, break the seal and insert a thin, rod-like small screwdriver
through the calibration hole and press the button as illustrated:
To change settings using the DIVA display follow the directions below.
At the beginning of the set up procedure the display will start off by showing the following.
Price Computing Version Weigh Only Version
Press the Zero Key:
Conf
Conf
1
2
Calibration Button
Thin Rod
Calibration button hole
is located underneath
the security label
1. Insert a thin rod into the hole directly underneath the seal
2. Move the thin rod toward the level bubble on the scale until
you feel the button depress