METTLER TOLEDO DIVA Service Manual PSC 8500 Magellan Scanner 72191202 (05/2012) 0.
INTRODUCTION This publication is provided solely as a guide for individuals who have purchased the METTLER TOLEDO DIVA scale product.
Precautions WARNING ONLY PERMIT QUALIFIED PERSONNEL TO SERVICE THIS EQUIPMENT. EXERCISE CARE WHEN MAKING CHECKS, TESTS AND ADJUSTMENTS THAT MUST BE MADE WITH POWER ON. FAILING TO OBSERVE THESE PRECAUTIONS CAN RESULT IN BODILY HARM. WARNING FOR CONTINUED PROTECTION AGAINST SHOCK HAZARD CONNECT TO PROPERLY GROUNDED OUTLET ONLY. DO NOT REMOVE THE GROUND PRONG. WARNING DISCONNECT ALL POWER TO THIS UNIT BEFORE REMOVING THE FUSE OR SERVICING. FAILURE TO DO SO MAY RESULT IN BODILY INJURY OR PROPERTY DAMAGE.
Contents 1 INTRODUCTION ........................................................................................................................................................................... 1 1.1 1.2 1.3 1.4 1.5 1.6 1.7 2 INSTALLATION ....................................................................................................................................................................... 6 2.1 2.2 2.3 2.4 2.5 3 PRECAUTIONS ................................................................
.1.11 6.1.12 6.1.13 ICL .............................................................................................................................................................................61 Shekel (WO) ...............................................................................................................................................................63 RIVA 5462 (WO) .........................................................................................................................
1 Introduction The DIVA is designed to mount on the PSC 8500 Scanner and communicate with ECR and POS systems for weighing needs. 1.1 Reliability The DIVA is developed, produced, and tested in a Mettler Toledo facility that has been audited and registered according to international ISO 9001 quality standards and ISO 14000 environment control program. 1.2 Standard Features 1 Capacity: 6x0.002kg;15 x 0.005kg – dual-interval 15x0.005lb; 30x0.
METTLER TOLEDO 1.
METTLER TOLEDO 1.4 Power Requirements Two different power options exist: 1. If the scanner is equipped for use with MT weighing modules, power will be supplied via a white FFC cable located in the front of the scanner. The DIVA-P2 is equipped with an appropropriate FFC socket for this cable. With this option, the scale is powered by the scanner’s power supply. 2.
METTLER TOLEDO Mettler-Toledo (ChangZhou) Scale & System Ltd. Legal Metrology Declaration of Conformity Déclaration de conformité Declaración de Conformidad Conformiteitsverklaring Dichiarazione di conformità We/Wir/Nous/Nosotros/Noi Mettler-Toledo (ChangZhou) Scale & System Ltd. 111 ChangXi Road, ChangZhou, JiangSu, 213001, P.R.
METTLER TOLEDO Important notice for verified weighing instruments M M M M Weighing Instruments verified at the place of manufacture bear the preceding mark on the packing label and a green “M” sticker on the descriptive plate. They may be set to work immediately. Weighing Instruments which are verified in two steps have no green “M” on the descriptive plate and bear the preceding identification mark on the packing label.
METTLER TOLEDO 2 Installation 2.1 Precautions Before you install your DIVA scale, keep in mind the following factors, which might have a negative influence on the scale's operation: Vibration diminishes the scale’s ability to measure accurately. Excessive vibration from equipment such as conveyors can cause inaccurate and non-repeatable readings. Be sure the scale is leveled properly. Air currents can also diminish a scales performance.
METTLER TOLEDO For Weigh Only Version: The Weigh Only version comes with a tower mount. Assemble the Weight Only tower as is illustrated here: Display Display Cable Display Bracket Display Tower Please refer to the Installation Guide for instructions on mounting the DIVA-P2 to the scanner. 1. Level the scale by turning the adjustable feet on the bottom of the unit. Adjust the feet until the level bubble is in the center of the circle. Bubble indicator: Correct Wrong 2.
METTLER TOLEDO 2.5 Sealing After installation is complete, legal-for-trade applications require sealing so that the settings cannot be changed. The label sealing sequence Security Sticker 28x19mm, material Vinyl white 3690 E UL, text color black. There are a total of three sealing labels, two of which are factory set. The third sealing label is needed to seal the calibration button and can be placed only by authorized personnel.
METTLER TOLEDO For your notes: DIVA Service Manual 9
METTLER TOLEDO 3 Set-up and Calibration There are two types of setup modes. The first is called “Service Mode” and it allows full access to all soft switches and to calibrate the scale. The second is called “Master Mode” and it only allows access to soft switches that do not affect the metrology of the scale. 3.
METTLER TOLEDO Price Computing Version Weigh Only Version Grp 1 Grp 1 You can press the Tare key to move to Grp 2, 3, or 4. To go into a particular Group, press the Zero Key. Below is what you will see when going into the first Step of Group 1. Price Computing Version Weigh Only Version Grp 1 Step 1 DE 1.1 DE Refer to Section 3.3 Setup and Soft Switch Default Table for all available settings. 3.
METTLER TOLEDO 3.3 Set up Soft Switch and Default Table Group.Step Function Possible Selections Master Mode Access Press Zero to Enter Press Tare to go to Group 2 Group 1 1.1 Country Select DE (Germany) FR (France) US (USA) CA (Canada) AT (Austria and Croatia and Slovenia) ES (Portugal and Spain) BE (Belgium, Netherland, Poland and England) IT (Italy) CH (Kazachstan and Switzerland) UA (Ukraine) RU (Russia) HU (Hungary) SK (Slovakia) CZ (Czechia) EXPORT (General export) CN (China) 1.
METTLER TOLEDO Press Zero to Enter Press Tare to go to Group 3 Group 2 Group.Step Function Possible Selections Master Mode Access 2.1 Tare Enable ON – Enables tare function. OFF – Disables tare function. Yes 2.2 Chain Tare Enable ON – Enables multiple tares. OFF – Only one tare per transaction is allowed. Yes 2.3* Total Price Round (To Nearest 0 or 5) ON – The total price will round up or down to 0 or 5. OFF – Total price will not be rounded. No 2.
METTLER TOLEDO Press Zero to Enter Press Tare to go to Group 1 EXIT SAVE Save or abort setting * Used only on the 4 Line display PC – Price Computing WO – Weight Only 14 DIVA Service Manual SAVE – save all settings and return to weighing mode ABORT – abort all settings and return to weighing mode Yes
DIVA Service Manual Currency 1.12 Off Flash Mode Enable Parity Data Bits Stop Bit Protocol 3.2 3.3 3.4 3.5 Weight Change Return to Zero 4.1 4.2 Group 4 Baud Rate 3.1 Group 3 2.9 2 Display Update Rate 2.
METTLER TOLEDO 3.
METTLER TOLEDO 3.
METTLER TOLEDO 18 DIVA Service Manual
METTLER TOLEDO 3.7 Calibration Note: These functions are directly related to the Weight & Measurement regulations in your country. Therefore they are protected by a sealed calibration button. The button is located as shown below. For the scale to be usable in a legal for trade application, the button must be sealed as is shown in section 2 after the calibration is complete.
METTLER TOLEDO Note: If the scale is unable to complete calibration it will display “E power to the scale and restart at Step 1 above. 20”. Cycle 12. The scale will now save the calibration and cycle out of the Service mode and return to normal weighing mode. 13. Verify the calibration was successful with the calibrated weight. calibration was not successful, restart at Step 1. If for some reason the 14. If applicable, you should remove power from the scale and apply the appropriate calibration seal.
METTLER TOLEDO 4 Operating Instructions 4.
METTLER TOLEDO 4.2 Keyboard Key Description Used to return the scale to gross zero. Tare key. 4.3 Cursors Cursor NET ZERO NET ZERO 22 DIVA Service Manual Description When tare is entered. When weight is gross zero.
METTLER TOLEDO 4.4 Operations 4.4.1 Weighing and Communication Place the item to be weighed on the platter, the weight will be displayed. Remote ASCII commands can control the scale through the provided RS232 port. Commands include zero, tare, and other data functions depending on the protocol. See Section 6 for available protocols and their descriptions. 4.4.2 Backlight function DIVA is equipped with a backlight for the display.
METTLER TOLEDO 5 Service and Maintenance This section provides information on servicing and maintaining the scale including: Cleaning and regular maintenance Troubleshooting WARNING ONLY PERMIT QUALIFIED PERSONNEL TO SERVICE THIS EQUIPMENT. EXERCISE CARE WHEN MAKING CHECKS, TESTS AND ADJUSTMENTS THAT MUST BE MADE WITH POWER ON. FAILING TO OBSERVE THESE PRECAUTIONS CAN RESULT IN BODILY HARM. WARNING DISCONNECT ALL POWER TO THIS UNIT BEFORE SERVICING OR CLEANING. 5.
METTLER TOLEDO 5.2 Troubleshooting The following table lists error messages, descriptions, and corrective actions. Error Codes and Actions E10 Generic error E 11 RAM error E 16 ROM error E 18 EEPROM error E 20 Calibration error Cycle power to the scale and restart calibration. E 28 Transmission error Reset scale. nnnnnn In weight display Over capacity nnnnnn in total price display Over 9999.99 Remove weight from Platter. If this does not work, try cycling the power to the scale.
METTLER TOLEDO Note: All other pins should not be connected. Should you need to construct your own communications cable. Dual 5 Connector directly on the scale. The following is the pinout on the Dual 5 Connector Pin Descriptions and Positions (when looking directly at the pins) Pin 9 Status Pin 7 RTS Pin 5 TxD Pin 3 +5V DC Pin 1 +12V DC Pin 10 Scale Zero Pin 8 CTS Pin 6 RxD Pin 4 Gnd Pin 2 Gnd 5.4.
METTLER TOLEDO 5.
METTLER TOLEDO Diva P3 Consists of same parts as DIVA P2 with the exception of the parts labeled below 28 DIVA Service Manual
METTLER TOLEDO Note: Most parts are not available for individual replacement. diagrams are only provided for reference.
METTLER TOLEDO 5.6 Replacement Parts Order Numbers Part Number Exploded ID 72191106 23-30 SL DISP,W/LENS&CABLE,MT,24P 72191107 31-52 4 Line, MT, €/KG, 6/15x.002/.005KG,15P 72187468 72187467 25 52 Lens, MT, 1 Line, 30 x .01LB Lens, MT, 1 Line, 6/15 x .002/.
METTLER TOLEDO Part Number Exploded ID 68001451 21 72194083 Not Shown 72185768 31 Currency Sticker,Poland, 4 Line Display 72185767 31 Currency Sticker,Czech, 4 Line Display 72185766 31 Currency Sticker,Hungary, 4 Line Display 71209973 31 Currency Sticker,Switzerland, 4 Line Display 71209974 31 Currency Sticker,Venezuela, 4 Line Display 71209972 31 Currency Sticker, $/kg, 4 Line Display 71209975 31 Currency Sticker,UK,kg, 4 Line Display 72190532 72190533 72188623 32 32 25 overlay
METTLER TOLEDO Part Number Exploded ID 71209952 34 32 DIVA Service Manual Description PCB, FOUR LINE DISPLAY Picture
METTLER TOLEDO 6 Appendix 6.1 Protocols The numbers correspond to the number that will show up in the Soft Switch settings in Group 3.5: 0. 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15.
METTLER TOLEDO 6.1.1 NCI – ECR 6.1.1.1 OVERVIEW The serial communications protocol is used to interface Weigh-Tronix/NCI POS (Point-of-Sale) scales to ECR (Electronic Cash Registers), or other computing equipment (eg. personal computers). Baud rate and parity will be configurable. Start and stop bits will each be fixed at one (1). Data bits will be fixed at seven. Baud rates supported will be 1200, 2400, 4800, 9600 and 19200.
METTLER TOLEDO Name: Command: Response: Zero the scale Z Scale is zeroed, returns scale status. (see note 4) Shhhh Optional Commands Name: Command: Response: Request high-resolution weight H Returns decimal wt in 10x with units plus scale status. (see notes 2 and 3) xxx.xxxUUShhhh Returns lb-oz wt in10x with units plus scale status. xLBxx.xxOZShhhh Scale status only if wt < 0, initial zero error, in motion or out of capacity.
METTLER TOLEDO Weight Command: W This command causes the scale to return the displayed weight string in ECR format. The scale will usually return the displayed weight with leading zeroes (ie no leading zero suppression). However, if weight is negative, in motion, over capacity or under capacity, or if a zero error exists, the ECR protocol causes only the scale's status to be returned in response to a weight ('W') command. The returned weight string will include decimal point plus units of measure.
METTLER TOLEDO Scale Status Command: S There will be at least two status bytes. If bit 6 of the second status byte is set then there will be a third byte. Bit 6 of each successive byte will indicate whether or not another status byte follows. The bit will be clear (0) in the last status byte. At this time only the first three bytes are defined. Others may be added in the future. Bit 0 is the least significant bit in the byte while bit 7 is the most significant bit.
METTLER TOLEDO 6.1.2 (Reserved) 6.1.3 8217 Mettler Toledo 6.1.3.1 Host Communications The 8217 scale acts as a peripheral device when connected to a host computer. When the host requests weight data by sending an uppercase W, the scale will respond with the weight data or a status byte if the scale is in motion or an invalid state.
METTLER TOLEDO 6.1.3.2 Command Descriptions ASCII Commands (* MUST BE UPPERCASE CHARACTERS ONLY.) *ASCII Command From Host Scale Response Description Instructs scale to send weight data. (W=Weight Digit 0-9). The scale will respond as follows according to setup and status: C STX WW.WW R STX WW.WWW CR C STX WW.WWN R STX WW.WWWN CR C STX?status byte R Gross Weight Pounds. Gross Weight KG. Net Weight Pounds. ASCII N added after weight. Net Weight KG. ASCII N added after weight.
METTLER TOLEDO 6.1.3.3 Status Byte Format When the scale responds with STX?status byte CR , this indicates a status byte. The status byte contains scale status information, which can include net/gross mode, zero status, etc. The actual status byte sent will be an ASCII character which must be converted to binary to decode the bits. The scale status bit definitions are shown in the following table.
METTLER TOLEDO 6.1.3.4 Confidence Test Status Byte Format When the scale responds with STX? status byte CR, after requesting a confidence test (B), a confidence test status byte will be sent. The status byte contains scale status information, which can include net/gross mode, zero status, etc. The actual status byte sent will be an ASCII character which must be converted to binary to decode the bits. The confidence test status bits are shown in the following table.
METTLER TOLEDO 6.1.3.5 Tare Tare can be taken on an item two ways: Automatic or digital. REMOTE TARE The unknown weight of an object can be tared-off automatically by placing the object on the scale platter, then with the display in a stable non-motion and positive weight condition, sending an uppercase ASCII T character followed by CR from the host will automatically subtract the weight. The tare function must be enabled. Chain tare is not permitted.
METTLER TOLEDO 6.1.4 8213 Mettler Toledo 6.1.4.1 Host Communications The 8217 scale acts as a peripheral device when connected to a host computer. When the host requests weight data by sending an uppercase W, the scale will respond with the weight data or a status byte if the scale is in motion or an invalid state.
METTLER TOLEDO 6.1.4.2 Command Descriptions ASCII Commands (* MUST BE UPPERCASE CHARACTERS ONLY.) *ASCII Command From Host Scale Response Description Instructs scale to send weight data. (W=Weight Digit 0-9). The scale will respond as follows according to setup and status: C STX 0WW.WW R STX WW.WWW CR C STX 0WW.WWN R STX WW.WWWN CR C STX?status byte R Gross Weight Pounds. Gross Weight KG. Net Weight Pounds. ASCII N added after weight. Net Weight KG. ASCII N added after weight.
METTLER TOLEDO 6.1.4.3 Status Byte Format When the scale responds with STX?status byte CR , this indicates a status byte. The status byte contains scale status information, which can include net/gross mode, zero status, etc. The actual status byte sent will be an ASCII character which must be converted to binary to decode the bits. The scale status bit definitions are shown in the following table. Scale Status Byte Bit Definitions Bit Number Bit Description DIVA Service Manual 7 7 Parity Bit (even).
METTLER TOLEDO 6.1.4.4 Confidence Test Status Byte Format When the scale responds with STX? status byte CR, after requesting a confidence test (B), a confidence test status byte will be sent. The status byte contains scale status information, which can include net/gross mode, zero status, etc. The actual status byte sent will be an ASCII character which must be converted to binary to decode the bits. The confidence test status bits are shown in the following table.
METTLER TOLEDO 6.1.5 EPOS 1 6.1.5.1 Data Format 2400 Baud (select 2400, 4800 or 9600) 7 data bits (select 7 or 8) Even parity (select Even, None or Odd parity) 1 stop bit (select 1 or 2 stop bit) 6.1.5.
METTLER TOLEDO 6.1.5.3 Command Flow ECR Scale COMMENTS ENQ (05h) ACK (06h) Data Available. Errors could be: CAN - repeat weighing NAK - no acknowledgement NUL - no data available Data Request Errors could be: NAK – no acknowledgement DC1 (11h) STX ID W5W4W3W2W1 BCC ETX Transmitted Data Confirm Data Errors could be: ACK – Data not confirmed STX ID W5W4W3W2W1 BCC ETX CR (0dh) Data confirmed Note: Spaces depicted in the description above are only used for ease of reading.
METTLER TOLEDO 6.1.6 EPOS 2 6.1.6.1 Data Format 2400 Baud (select 2400, 4800 or 9600) 7 data bits (select 7 or 8) Even parity (select Even, None or Odd parity) 1 stop bit (select 1 or 2 stop bit) 6.1.6.2 Handshaking For PC DB25 serial port connector: Jumper 4 to 5, and 6, 8, & 20 together. For PC DB9 serial port connector: Jumper 4 to 6 and 7 to 8. 6.1.6.
METTLER TOLEDO 6.1.6.4 Command Flow ECR Scale COMMENTS ENQ (05h) ACK (06h) Data Available. Errors could be: CAN - repeat weighing NAK - no acknowledgement NUL - no data available Data Request Errors could be: NAK – no acknowledgement DC1 (11h) STX ID W5W4W3W2W1 BCC ETX Transmitted Data Note: Spaces depicted in the description above are only used for ease of reading. No space characters are used unless the BCC result yields the space character.
METTLER TOLEDO 6.1.
METTLER TOLEDO 6.1.8 Dialog 06 With free programmable POS-Systems it is for a third party on principle possible, to manipulate parts of the software, which are obligate to verification. For this reason certifying agencies expects appropriate protection against such inadmissible manipulations of third parties. A protection against manipulations presents the Checkout-Dialogue 06 in conjunction with precautions taken at the POS-Software. 6.1.8.
METTLER TOLEDO 6.1.8.3 Command Descriptions Records for Communications from a POS system to the Scale Record 01: Transmitting of unit price EOT STX 01 ESC D5 D4 D3 D2 D1 D0 ESC ETX \_/ Record No. Record 03: \_______________________/ Unit price 5/6 digits Transmitting of unit price and tare value EOT STX 03 ESC D5 D4 D3 D2 D1 D0 ESC T3 \_/ Record No.
METTLER TOLEDO Records for Communications from the Scale to a POS system Record 02: Result of weighing (answer to records 01, 03, 04, 05) STX 02 ESC X ESC D4 D3 D2 D1 D0 ESC D5 D4 D3 D2 D1 D0 ESC D5 D4 D3 D2 D1 D0 ETX \_/ Record No. I \_______________/ scale-weight 5 digits I status \__________________/ unit price 5/6 digits \__________________/ price to pay 6 digits Scale status 30H = lb. : oz / 1/8 oz 31H = lb. / 0,01 32H = lb.
METTLER TOLEDO 6.1.8.4 Description of the Dialog A dialogue will always be initiated by the POS-system. Description of weighing operation The POS-system transmits one of the Records 01, 03, 04 or 05. The scale answers (if no errors) with ACK and displays the transmitted data. The POS-system now can request the weighing result from the scale by transmitting EOT ENQ. If the result is known the scale answers with Record 02. If there is an error, the scale answers with NAK.
METTLER TOLEDO 6.1.8.5 Flow Chart STX EOT STX SATZ 01 ETX SATZ 11 (32H, Z) ETX ACK EOT NAK STX SATZ 03 ETX STX SATZ 04 ETX STX SATZ 05 ETX STX SATZ 10 ETX STX SATZ 20 ETX ENQ STX SATZ 02 ETX STX SATZ 11 (31H ) ETX STX SATZ 11 (30H ) ETX NAK STX STX SATZ 08 ETX NAK STX SATZ 80 ETX ACK STX SATZ 81 ETX NAK Kasse Waag e 6.1.8.
METTLER TOLEDO 6.1.9 Dialog 04 / 02 Note: To use Dialog 04, select this protocol but set the baud rate to 4800, for Dialog 02 the baud rate is set to 2400 6.1.9.1 Data Format Baud rate: 2400 Baud Type of transmission: asynchronous Data format: 7 Bit + Parity Parity: odd Stop bits: 1 Bit Connection Scale TXD 3 RXD 2 GND 5 6.1.9.
METTLER TOLEDO 6.1.9.
METTLER TOLEDO Scale Communications to a POS System Setting number 02: Weight, Price, and Total Price STX 30h 32h ESC X ESC D4 D3 D2 D1 D0 ESC \____/ setting no. (02) I \_____________/ I scale - weight status 5 digits 30H = lb 31H = lb. / 0,01 32H = lb. / 0,005 33H = kg D5 D4 D3 D2 D1 D0 ESC \________________/ Basic price 6 digits D5 D4 D3 D2 D1 D0 ETX \________________/ Price to pay 6 digits Setting number 09: Status information STX 30h 39h ESC S1 S0 ETX \____/ setting no.
METTLER TOLEDO 6.1.10 Extended Dialog 06 The Extended Dialog 06 protocol is for use with communication between the DIVA and the METTLER TOLEDO VCODisp module.
METTLER TOLEDO 6.1.11 ICL 6.1.11.1 Data Format 9600 Baud (select 2400, 4800 or 9600) 7 data bits (select 7 or 8) Even parity (select Even, None or Odd parity) 1 stop bit (select 1 or 2 stop bit) 6.1.11.2 Handshaking For PC DB25 serial port connector: Jumper 4 to 5, and 6, 8, & 20 together. For PC DB9 serial port connector: Jumper 4 to 6 and 7 to 8. 6.1.11.
METTLER TOLEDO 6.1.11.4 Command Flow ECR Scale COMMENTS ENQ (05h) ACK (06h) Data Available. Errors could be: CAN - repeat weighing NAK - no acknowledgement NUL - no data available Data Request Errors could be: NAK – no acknowledgement DC1 (11h) STX ID W5W4W3W2W1 BCC ETX Transmitted Data Confirm Data Errors could be: ACK – Data not confirmed NAK – receive error or scale error detected.
METTLER TOLEDO 6.1.12 Shekel (WO) 6.1.12.1 Command Flow ECR S Scale S_XXXXXXXXXX_g_ SI_ Comment Scale sends valid (>= 0, no overload) and stable weight Sent if no valid and stable weight available for 2 sec.
METTLER TOLEDO 6.1.13 RIVA 5462 (WO) 6.1.13.
METTLER TOLEDO 6.1.14 IP3 (PC) 6.1.14.
METTLER TOLEDO 6.1.14.2 ECR Command Descriptions Time out Scale COMMENTS STX (02h) 500ms ACK (06h) @ (40h) 1 (31h) P4 (30h to 39h) P3 “ “ P2 “ “ P1 “ “ P0 “ “ CR (ODh) LF (0Ah) The cash register sends the unit price (P1-4) to the scale. When received, the scales send back the weight, the unit price and the price to pay. All information are memorized in the cash register (PLU, Totals etc..).
METTLER TOLEDO 6.1.
METTLER TOLEDO METTLER TOLEDO 1900 Polaris Parkway Columbus, Ohio 43420-2020 P/N: 72191204 METTLER TOLEDO® is a Trademark of Mettler-Toledo, LLC ©2012 Mettler-Toledo, LLC 68 DIVA Service Manual