Installation and Operation Guide

Table Of Contents
60
VIBCONNECT RF 05.2012
5.6.4 Creating measuring configuration
This chapter explains how to create a measuring configuration for
a measurement location in the VIBCONNECT RF system, using the
OMNITREND software. A measuring configuration contains infor-
mation regarding
the measuring tasks
the time intervals at which measurements are to be performed
the measurement location assigned to a sensor unit
the associated bridge.
Measuring tasks
Select a measurement location.
Note
Always observe the channel assignment:
Channels 1 and 2 are reserved for vibration measurements;
channels 3 and 4 are reserved for temperature measurements
(see also '5.4.4 Connecting sensor unit', page 45).
In the main menu, select <Machinery Manager> / <Add> /
<Measuring Task>.
In the measuring task dialog, activate option <Select Setup
Directly>.
Select the applicable measuring task and click the <Next> button.
In the next window, click the <Create> button.
If you wish to set up another vibration measuring task, repeat
the above steps (max. 4 tasks per measurement location).
For the calculation of the characteristic vibration values, define
the respective band analysis in the alarm wizard.
Installation and commissioning
The characteristic values
are calculated from
a spectrum by means
of “band analyses”