Installation and Operation Guide
Table Of Contents
- Chapter 1: Introduction
- 1.1 First steps
- 1.2 Service addresses
- 1.3 About this manual
- Chapter 2: Safety
- 2.1 Safety symbols
- 2.2 Information for the system operator
- 2.3 Information for operating personnel
- 2.4 Intended use
- 2.5 Residual risks and safety measures
- Declaration of conformity
- Certifications
- Chapter 3: Technical data
- Chapter 4: System description
- 4.1 VIBCONNECT RF bridge
- 4.2 VIBCONNECT RF sensor unit
- 4.3 VIBCONNECT RF sensor
- Chapter 5: Installation & commissioning
- 5.1 Quick guide
- 5.2 Configuration in OMNITREND
- 5.3 Installation & commissioning of bridge
- 5.4 Installation & commissioning of sensor unit
- 5.5 Installation of sensors
- 5.6 Configuration in OMNITREND (continued)
- 5.7 Installing additional sensor unit
- Chapter 6: Maintenance
- 6.1 Cleaning
- 6.2 System time correction
- 6.3 Installation report
- 6.4 Battery-powered sensor unit
- 6.5 Update
- 6.6 Warranty
- 6.7 Spare parts and accessories
- Chapter 7: Troubleshooting
- Chapter 8: After use
52
VIBCONNECT RF 05.2012
Mount the adapter:
• Press the adapter with minimum force against the mounting
area and turn it slightly to distribute the adhesive evenly.
• Do not remove adhesive escaping at the side of the adapter. For
extra stability, also apply adhesive around the adapter edge.
• Allow the adhesive to cure (approx. 24 h at room temperature).
Note
If required, secure the adapter during the first hour of the curing
period with the magnetic gluing aid VIB 8.477 or adhesive tape.
• Check the adapter for proper fixture (adapter must not rock).
• Screw the sensor onto the adapter and tighten it with a size 19
socket wrench. Permissible tightening torque: 10 – 20 Nm!
Mounting sensor with magnetic adapter
Note
The magnetic adapter is obviously only suitable for attachment to
ferromagnetic parts with adequate iron content. This mounting
option is only recommended if screw mounting or attachment
with adhesive adapter is not possible.
• Screw the magnetic adapter VIB 3.423 to the sensor and attach
it to the measurement location.
Installation and commissioning
Apply adhesive Press down and
turn adapter