Check the power indicating LED, and then review firmware version of PRUDENT in
the Config folder of Sensorprobe.
- LAN cable
Check unidentified network at network connection in the control panel.
- USB cable
Proceed to the control panel, device Manager and check the Virtual Serial Port
Should you have unidentified devices, be sure all cables are connected tightly. See
Chapter. 3.2. and Chapter. 5. for more information.
If you cannot solve the problems, contact the technical support team.
Have you performed initial Sensorprobe set-up?
Status
Sensorprobe should be set up at first in order to link with acquisition program.
Tips
Check if PRUDENT is properly inked with the acquisition program.
If not linked, see Chapter. 4-2 Sensorprobe, and set up Sensorprobe.
Contact the technical support team, if you cannot solve the problems.