Users Manual

Install & Service Manual AT-SM-04
Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8
2020. PIXXGEN Corporation. All rights reserved. -68-
Appendix - Check List
Is PICBs power on?
Status
PRUDENT is powered by PICB.
Tips
Check the Blue light at the front of PICB.
If you still have the light turned off even though all problems are solved, consult with
our friendly technical support team.
Have you installed Gigabit NETWORK?
Status
PRUDENT demands gigabit communication for fast image transmission. It is essential
to install gigabit network to operate PRUDENT system.
Tips
Check network driver version and network speed at network connection in the control
panel.
If the speed of network is not faster than 1 gigabit, see Chapter. 3-1 Network Adapter
Setting.
Should you still have lower speed than 1 gigabit, please contact the technical support
team.
Are all cables connected tightly?
Status
It is impossible to obtain clear images unless the calibration data saved correctly.
Tips
Check the calibration data in C://DR_data/Calibration data folder.
If no data exists in the folder, create one.
See Chapter. 8 Calibration for more information.
Are all cables connected tightly?
Status
Be sure all the connecting cables are secured tightly.
- HDMI cable
Power supplying and data transmitting to PRUDENT
- LAN cable
Sending image to acquisition program
- USB cable
Connecting devices for manual mode
Tips
- HDMI cable