Users Manual
Install & Service Manual AT-SM-04
Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8
ⓒ 2020. PIXXGEN Corporation. All rights reserved. -56-
In any case the study will be less than 4 seconds.
1. The x-ray technique should be the same or less than the traditional study. This is because
there is persistence in the scintillator and during longer exposures this can be taken
advantage of by decreasing mA. Experience will show the technologist how much to
decrease the mA.
2. Once the time to go from the start position to the finish is determined it should be set on
the generator control.
3. The frames per second should be determined by the physician based on how many views
he believes are needed.
4. Then the technologist will position the patient in the “start” or first position. Have the
patient move the body part to the second or “End” position at a comfortable speed. This
should be repeated a couple of times.
5. Return the patient to the “Start” position and obtain the radiograph normally.