CAUTION!
DO NOT “MAKE BPM” YOURSELF.
Do not use MAKE BPM function after calibration. BPM file is provided.
After calibration, copy BPM file from PRUDENT USB to the following folder –
“C: //DR_Data/Calibration”.
In case the user performs MAKE BPM, there might be the deterioration of
image quality.
Tip!.
If there is a visible line on the image, performing 5-minute calibration is
recommended.
Turn off PRUDENT and turn it on again after 1 hour.
After previous process, perform 5-minute calibration after 5 minutes.
The visible line on the image will be gone if the 5-minute calibration data is in
use.