Users Manual

Install & Service Manual AT-SM-04
Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8
2020. PIXXGEN Corporation. All rights reserved. -49-
8.2.2. GET BRIGHT
Bright file will be named as Bright_xxxxx after GET BRIGHT process.
Bright value is proportional to the strength of X-ray (kV) and exposure time (mAs or sec).
Only one bright image is required for the calibration data of PRUDENT.
Delete unnecessary files by clicking ‘DELETE FILE’
*Click [GET BRIGHT] button and exposure and X-ray after 2seconds