Users Manual
Install & Service Manual AT-SM-04
Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8
ⓒ 2020. PIXXGEN Corporation. All rights reserved. -46-
8. Calibration
8.1. Calibration Data Installation
CAUTION!
Allow at least 30 minutes to warm up for the maximum performance of
instrument.
PRUDENT requires calibration data for best quality imaging. An optimal calibration data for
PRUDENT has been provided on the PRUDENT USB. In case of lost or broken PRUDENT USB,
however, the user may perform calibration with Sensorprobe.
Insert PRUDENT USB.
Copy ‘Calibration Data’ folder.
Create ‘C: //DR_data/’ folder, and paste ‘Calibration data’ folder into the ‘DR_data’ folder.
The calibration data destination folder varies depending on what acquisition program the user
utilizes. Check the destination folder first when creating calibration data.
CAUTION!
Both ‘SENSOR.ini’ file and the calibration data must be saved in the same
folder. Otherwise, PRUDENT will not be able to read the calibration data. For
more information for ‘SENSOR.ini’ file, see Appendix: ‘SENSOR.ini’ file setting
Tip!.
These calibration data creation instructions are not necessary for the user who
uses the calibration data provided.
CAUTION!
If there is no calibration data in the designated destination, the warning
message above will show up.