6.2.1. Installation for Station Mode
or
⚫
Bridge ⇄ Work Station: Wired or Wireless (signal, digital image)
⚫
Place the Bridge on the wall in the middle of Stand and Table Bucky.
⚫
Elevate the Bridge away from floors. (Proper Height: To be horizontal with Table Bucky.)
⚫
Position the Bridge antenna vertically, so that the antenna is standing straight up.
CAUTION!
A disconnection of wireless signal can be easily occurred depends on the
location of antenna. It has to minimize the shadow zone by installing additional
extended cable, so the product and the antenna can be opposable.