Users Manual
Install & Service Manual AT-SM-04
Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8
ⓒ 2020. PIXXGEN Corporation. All rights reserved. -24-
4.1.2.3. Open Config
(1) Trigger Mode: The trigger mode of PRUDENT
1 – Auto Mode
11 – Continuous Auto Mode
6 – Manual Mode
66 – Continuous Manual Mode
(2) Multi Shot: The number of images acquired at a time
(3) Firmware: The firmware version of PRUDENT.
(4) Mac address: Media Access Control address of PRUDENT
(5) RDY Time(ms): Push ready button(Manual mode).
(6) EXP Time(ms): Push exposure button(Manual mode), exposure time(Auto mode)
(7) Battery: Battery remains
(8) Temperature : Temperature
(9) Read: Read detector parameters.
Write: Write detector parameters
Get log: Bring log information
RESET: Reset the detector
REBOOT: Reboot the detector
Date/Time Sync: Synchronize date and time
between detector and workstation
CAUTION!
It is strongly recommended that the user is not to modify the contents of
Sensorprobe Config except authorized engineers from PIXXGEN.
If setting is modified, it may result in malfunction.