3. Specifications
3.1. PRUDENT Specifications
Application
General or Dynamic Radiography
Sensor
IGZO or a-Si TFT array Flat Panel Detector
Conversion Screen
GOS / CsI
Active Area
Prudent1717: 17 X 17 Inch (430mm x 430mm)
Prudent1417: 14 X 17 Inch (350mm x 427mm)
Prudent1212: 11.3 X 11.3 Inch (287mm x 287mm)
Sensor Pixel
Prudent1717: 4302x4302 / 3072x3072 / 2560x2560
Prudent1417: 3534x4302/ 2500x3052
Prudent1212: 2867x2867/ 2048 X 2048
Image Data
16bit
Trigger Mode
AED (Automatic Exposure Detection),Manual Trigger
Calibration Mode
ACC (Automatic Calibration Control), Manual Calibration
Data Transfer Time
Less than 1 sec
Capture Cycle Time
2 ~ 5sec. (Approx.)
Wired interface
Gigabit Ethernet
Voltage
Adaptor
18 V d.c., 3.5 A (HPU63A-107)
Battery
14.8 V d.c., 3300 mAh
Battery Performance
Approx. 8 hours of continuous operation
Lithium ion Polymer Battery
Dimensions(W x L x T)
Prudent1717: 460mm(W) x 460mm(L) x 15mm(T)
Prudent1417: 382mm(W) x 460mm(L) x 15mm(T)
Prudent1212: 327mm(W) x 315mm(L) x 15mm(T)
(Cassette-size for the standard table or wall Bucky)
Weight
Prudent1717 : 3.9Kg ±10% (with battery)
Prudent1417 : 3.3Kg ±10% (with battery)
Prudent1212 : 2.0 Kg ±10% (with battery)
3.2. Environment
Operating condition
Temperature (C)
10C - 35C
Relative humidity (%)
20% - 75%
Atmospheric Pressure (hPa)
700hPa - 1060hPa
Transport and storage
condition
Temperature (C)
-20C 70C
Relative humidity (%)
10% - 95%
Atmospheric Pressure (hPa)
400hPa - 1500hPa