1.6. Condition for installation
Do not install this equipment in any of the locations listed below.
- Where the power supply is unstable.
- Where temperature and humidity are high.
- Where the room is without air-conditioner or ventilation.
- Where it is exposed to direct sunlight.
It is highly recommended to separate X-ray power in the distributing board of the electricity
of the building.
It is highly recommended to take internet connection nearby workstation pc in the room
Check the distance between workstation PC in the operation room and the area which
detector will be placed.
The wall outlet or the circuit breaker shall be installed near the equipment and shall be
easily accessible when problems occur.
Turn off the detector after the operation.
- For wired Detector: Press the stand-by button and check the power display
(Green ->Red)
- For wireless Detector: Press the Stand-by button and remove the battery pack. (Lift up
the battery pack and remove it.)
1.7. Environment
Be sure to use and store this equipment under the conditions described below.
Temperature
Humidity
Storage
-20 to 70
10 to 95 % RH(Non-condensing)
Operation
10 to 35
20 to 75% RH(Non-condensing)
Do not expose this equipment to high temperature and humidity since it makes detector
malfunctions.
1.8. Etc.
Type of protection against electric shock
- Class I and/or Internally powered equipment.
Degree of protection against electric shock
- Not classified - No applied parts
Classification according to the degree of protection against ingress of water
- IPX0, ordinary equipment
This equipment is not suitable for use in the presence of flammable anesthetics or oxygen
Mode of operation:
- continuous operation