Users Manual
Install & Service Manual AT-SM-04
Prudent │ Flat-panel Digital X-ray Detector Rev.1.0.8
ⓒ 2020. PIXXGEN Corporation. All rights reserved. -13-
1.6. Condition for installation
• Do not install this equipment in any of the locations listed below.
- Where the power supply is unstable.
- Where temperature and humidity are high.
- Where the room is without air-conditioner or ventilation.
- Where it is exposed to direct sunlight.
• It is highly recommended to separate X-ray power in the distributing board of the electricity
of the building.
• It is highly recommended to take internet connection nearby workstation pc in the room
• Check the distance between workstation PC in the operation room and the area which
detector will be placed.
• The wall outlet or the circuit breaker shall be installed near the equipment and shall be
easily accessible when problems occur.
• Turn off the detector after the operation.
- For wired Detector: Press the stand-by button and check the power display
(Green ->Red)
- For wireless Detector: Press the Stand-by button and remove the battery pack. (Lift up
the battery pack and remove it.)
1.7. Environment
• Be sure to use and store this equipment under the conditions described below.
Temperature
Humidity
Storage
-20 to 70℃
10 to 95 % RH(Non-condensing)
Operation
10 to 35℃
20 to 75% RH(Non-condensing)
• Do not expose this equipment to high temperature and humidity since it makes detector
malfunctions.
1.8. Etc.
• Type of protection against electric shock
- Class I and/or Internally powered equipment.
• Degree of protection against electric shock
- Not classified - No applied parts
• Classification according to the degree of protection against ingress of water
- IPX0, ordinary equipment
• This equipment is not suitable for use in the presence of flammable anesthetics or oxygen
• Mode of operation:
- continuous operation