User's Manual
LBL-0151 | User Supplemental Information, Effective: xxXXX2013
5
Extended-Use PATCH
Guidance and manufacturer’s declaration –
electromagnetic immunity
The Proteus Patch is intended for use in the electromagnetic environment specified
below. The customer or the user of the Patch should ensure that it is used in such an
environment.
Immunity test IEC 60601 test level
Compliance
level
Electromagnetic
environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/- 6kV contact
+/- 8kV air
+/-
6kV
contact
+/- 8kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast
transient / burst
IEC 61000-4-4
+/- 2 kV for power
supply lines
+/- 1 kV for
input/output
lines
Not
applicable
Surge IEC 61000-
4-5
+/- 1 kV line(s) to
line(s)
+/- 2 kV line(s) to
earth
Not
applicable
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % UT (>95 % dip in
UT)
for 0,5 cycle
40 % UT (60 % dip in
UT)
for 5 cycles
70 % UT (30 % dip in
UT)
for 25 cycles
<5 % UT (>95 % dip in
UT)
for 5 s
Not
applicable
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3A/m 3A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.