User's Manual

LBL0176,Rev1|UserSupplementalInformation,
Proteus®Patch
Page5of5 Effective:19DEC2013
Guidanceandmanufacturer’sdeclarationelectromagneticimmunity
TheProteusPatchisintendedforuseintheelectromagneticenvironmentspecifiedbelow.The
customerortheuseroftheProteusPatchshouldassurethatitisusedinsuchanenvironment.
Immunitytest IEC60601
testlevel
Compliance
level
Electromagneticenvironment
guidance
PortableandmobileRF
communicationsequipmentshouldbe
Guidanceandmanufacturer’sdeclarationelectromagneticimmunity
The Proteus Patch is intended for use in the electromagnetic environment specified below. The
customerortheuserofthePatchshouldassurethatitisusedinsuchanenvironment.
Immunitytest IEC60601testlevel Compliance
level
Electromagneticenvironment
guidance
Electrostatic
discharge(ESD)IEC
6100042
+/‐6kVcontact
+/‐8kVair
+/‐6kV
contact
+/‐8kVair
Floorsshouldbewood,concrete
orceramictile.Iffloorsare
coveredwithsyntheticmaterial,
therelativehumidityshouldbeat
least30%.
Electricalfast
transient/burst
IEC6100044
+/‐2kVforpower
supplylines
+/‐1kVfor
input/output
lines
Notapplicable
SurgeIEC610004
5
+/‐1kVline(s)to
line(s)
+/‐2kVline(s)toearth
Notapplicable
Voltagedips,short
interruptionsand
voltagevariations
onpowersupply
inputlines
IEC61000411
<5%UT(>95%dip in
UT)
for0,5cycle
40%UT(60%dipinUT)
for5cycles
70%UT(30%dipinUT)
for25cycles
<5%
UT(>95%dipin
UT)
for5s
Notapplicable
Powerfrequency
(50/60Hz)
magneticfield
IEC6100048
3A/m 3A/m Powerfrequencymagneticfields
shouldbeatlevelscharacteristic
ofatypicallocationinatypical
commercialorhospital
environment.
NOTEUTisthea.c.mainsvoltagepriortoapplicationofthetestlevel.