User Manual

LBL-xxxx Page 8
Proteus Two-part Patch Technical description Effective: xxXXX2016
Guidance and manufacturer’s declaration
electromagnetic immunity
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % UT (>95 % dip in
UT)
for 0,5 cycle
40 % UT (60 % dip in
UT)
for 5 cycles
70 % UT (30 % dip in
UT)
for 25 cycles
<5 % UT (>95 % dip in
UT)
for 5 s
Not applicable
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30A/m 30A/m Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration
electromagnetic immunity
The Proteus Patch is intended for use in the electromagnetic environment specified below. The
customer or the user of the Proteus Patch should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic environment
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80
MHz
10V/m
80 MHz to 2,5
GHz
Not Applicable
10 V/m
Portable and mobile RF
communications equipment should be
used no closer to any part of the
Proteus Patch, including cables, than
the recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
Not Applicable
d = 1,17 P 80 MHz to 800 MHz
d = 2,33 P 800 MHz to 2,5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the