User Manual
LBL-xxxx Page 7
Proteus Two-part Patch Technical description Effective: xxXXX2016
antennas) should be used no closer than 30 cm (12 inches) to any part of the Patch. Otherwise,
degradation of the performance of the Patch could result.
Guidance and manu
f
acturer’s declaration
–
electromagnetic emissions
The Proteus Patch is intended for use in the electromagnetic environment specified below. The
customer or the user of Proteus Patch should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The Patch uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B The Patch is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Not applicable
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
Guidance and manufacturer’s declaration
–
electromagnetic immunity
The Proteus Patch is intended for use in the electromagnetic environment specified below. The
customer or the user of the Patch should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance
level
Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/- 8kV contact
+/- 15kV air
+/- 8kV
contact
+/- 15kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient / burst
IEC 61000-4-4
+/- 2 kV for power
supply lines
+/- 1 kV for
input/output
lines
Not applicable
Surge IEC 61000-
4-5
+/- 1 kV line(s) to
line(s)
+/- 2 kV line(s) to earth
Not applicable