Specifications
Troubleshooting 71
SCOTT SAFETY 087-0048 REV B
Test Station
Hardware
(Continued)
Run time below
requirements or
battery damaged
• Verify connection.
• Replace Lithium Rechargeable Battery, item #6 in
Figure 50.
Bump or Calibration
Push Button loss of
function or shows
physical damage
• Verify connection and function.
• Replace applicable Push Button Switch, item #35 in
Figure 51, and in Figure 52.
USB Cover Bracket
shows physical
damage or
misplaced
• Replace USB Cover Bracket, item #26 in Figure 51,
and Figure 52.
USB Shroud Block
shows physical
damage
• Replace USB Shroud Block, item #27 in Figure 51
and Figure 52.
Lithium Battery
Holder shows
physical damage
• Replace Lithium Battery Holder, item #7 in
Figure 50.
IR PCB Shroud
Block shows
physical damage
• Replace IR PCB Shroud Block, item #28 in
Figure 51 and Figure 52.
Gas leak • Replace Teflon tape.
• Replace 1/8” High Pressure Tube Fitting, item #33 in
Figure 51.
Tygon tubing does
not stay in place or
excess leaking
• Replace Bulkhead Fitting 5/16” Tube, item #32 in
Figure 51, and Figure 52.
Improper readings or
gauge leaks
• Replace Gauge, item #36 in Figure 51.
Calibration Cap
damaged
• Replace Calibration Cap Modification, item #2 in
Figure 50.
Calibration Cup
Holder Bracket
damaged
• Replace Calibration Cup Holder Bracket, item #3 in
Figure 50.
Calibration Bottle
Mount damaged
• Replace Calibration Bottle Mount Kit, item #29 in
Figure 51.
Table 24 Troubleshooting Matrix (continued)
ITEM SYMPTOM/CAUSE SOLUTION