Retrieving Event Logs
Using the Test Station
Software
This section describes retrieving the Monitors event log using the Test Station
Software.
Event logs are listed by serial number. The event log retrieved are up to the last twenty
five (25) most recent events. Past events can be located by accessing the USB pen
drive.
1 Select either IR or USB Disk and then click READ, depending on the location of
the USB Pen Drive.
2 All available monitors are listed by serial numbers. To select the file(s) you want,
check the box next to the desired monitor(s). You may download multiple event
logs at the same time. By default these files are .CSV and should be opened using
MS Excel. See Figure 46.
Figure 46 Test Station Software Event Logs
Individual event logs (eventxxx.csv) are organized inside the ProZM folder by the
serial number. The serial number format is MGYYWWxxxx (where M=Monroe code,
G=Gas type (H,O,C), YY=Year, WW=Week and xxxx is the 4-digit serial number).