Specifications

64 CHAPTER 6: MAINTENANCE
087-0048 REV B SCOTT SAFETY
Second line
These are unique
to each monitor
and appear for
each test
Bay Num 1,2,3,4 these correspond with the matching Unit numbers
on the test station
Detector Monitor or Monitor with Hibernation mode.
Serial Serial Number for each monitor
Result Pass/Fail
Monitor F/W
Ver si on
Monitor firmware version
Monitor H/W
Ver si on
Monitor hardware version
Audio Pass/Fail
Num Sensors Number of sensors in the monitor, this should always
display 1.
Options
(Show Sensor
Reading,
Bump Flash
Disabled and
Hide Clock)
T (True) if enabled; F (False) if disabled.
Options
(User ID)
Unique 6 digit number assigned by user.
Options
(Self Test
Interval)
Default is 20hours; this option may be changed via the IR
Connect or via the Test Station Software.
Third line
For each sensor
on the monitor
Sensor Senor Type (H
2
S, CO or O
2
)
Result Pass/Fail
Reading Peak unit reading during test in PPM or %
Bottle Lot# Field configured via Test Station Software.
Bottle Expiry Field configured via Test Station Software.
Bottle
Concentration
Default settings may be found and configured via the
Test Station Software.
Options
(High Alarm,
Low Alarm)
Alarm settings of the individual monitor.
Options
(Calibration
Interval and
Bump Test
Interval)
Intervals of the individual monitor.
Options
(Bump Test
Due and
Calibration
Test Due)
Number of days monitor scheduled test is due.
Table 21 Test Record Line by Line Details (continued)
LINE FIELD DESCRIPTION