Maintenance This section covers maintenance requirements.
Self-Testing the
Monitor
This section covers self-testing the monitor.
Prior to daily use, the device prompts to perform a self-test. This process is a simple
and effective way to ensure safe operation of the monitor. During the self-test, the
audio, visual and vibration alarms are activated and the sensor is tested. Table 20
details a step by step process for performing the self-test.
Warning: The Self-test does not take place of a Bump Test or a Calibration to ensure
the monitor response to gas.
Table 20 Self-Test Steps
LCD STEPS
When the TEST icon appears in the upper left hand corner, a self-test is
required. Press the button on the front of the monitor to perform the
self-test.
After pressing the button this screen appears. During the self-test ensure
the following occurs:
The monitor emits one audible beep
All LEDs light and monitor vibrates
All LCD display elements appear
After the full element LCD displays, the low alarm and high alarm set
points display.
Note: These alarm set points may be changed using either the IR Connect
Software or Test Station Software. See “Specifications” on page 80.
Note: Once a self-test is performed successfully, the Check Mark appears
automatically, and the High and Low set points display.
Note: Provided the monitor has not been programmed via the IR Connect
Software or has not been exposed to gas, this displays. Otherwise, go to
the next step.
When a self-test is successful the monitor turns to the original screen and
displays a CHECK MARK in place where the TEST icon was previously
displayed and one short audible beep sounds.
The monitor by default prompts another self-test in twenty (20) hours
from when the button was pressed.
Note: This value may be changed using the IR Connect Software.
(If applicable) If programmed with a USER ID, after the alarm set points
are displayed, a combination of numbers or letters scrolls across the
LCD. This includes a maximum of two (2) screens with a maximum
character limit on the USER ID of six (6) characters. The USER ID may
be changed via the IR Connect Software.