Specifications
36 CHAPTER 4: CONFIGURATION AND SETUP
087-0048 REV B SCOTT SAFETY
Configuring the Test
Station Software
Parameters
This section describes configurating the Test Station Software parameters.
The Test Station Software consists of the following three (3) Tabs:
• Test Station — Allows changes to Test Station parameters
• Protege ZM — Allows various changes to each sensor type. Once this
configuration has been sent to the Test Station, every monitor that is Bump Tested
or Calibrated takes on these new configuration parameters.
• Logs — Allows retrieving Test Records, Event Logs and Inventory Files. See
“Maintenance” on page 55.
Warning: Parameter changes written to the Test Station over ride those Monitor
parameters when Monitors are Bumped or Calibrated. Failure to follow these
instructions may result in serious injury or death.
When the Test Station reads a configuration file, it uses those settings until it shuts
down or until a new configuration file is inserted. If the Test Station is powered and
then the USEB Pen Drive is removed, it continues to use those settings. When the Test
Station is turned off and on again with no USB Pen Drive installed, then the Test
Station starts using the firmware defaults.
Figure 28 shows the Test Station Tab screen and Table 13 details the parameters.
These parameters allow you to program the Test Station to met your needs.
Figure 28 Test Station Software Read